New TopShim run:

version = topshim 4.4.0
date = 2024-09-05
time = 14:14:44.282
command = topshim hrmas

HRMAS SHIMMING

Parameters:

probe = Z180004_0001
angle PHI = 270.5 degree
solvent = H2O+D2O
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:

Iteration 1: FWHM Before= 1.0 Hz, Target= 0.2 Hz, After= 2.1 Hz
Iteration 2: FWHM Before= 2.1 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 3: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 4: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, MAS) = 0.469


MAS shims:
Z'	-15176
Z2'	+94351
Z3'	-55328
Z4'	+0


duration = 1 min 31 sec
completed successfully
finished Thu Sep  5 14:16:14 2024