start configuration on server
wait for server to start hardware configuration
get permission to start a new configuration
start a new configuration
reinitialize objects
determine instrument name
create directories
read old configuration
edit parameters
send user input to server
wait for server to handle parameters
get permission to continue configuration
continue configuration
parse input from user
check for questions from server
check hardware
there are 8 DHCP controlled devices and 18 devices with fixed IP to check
try to connect 26 devices at the spectrometer subnet
connected: CU5 - COOLING UNIT/5 Z159073/049 at IP 149.236.99.254
connected: BLA_W1345092_0382 at IP 149.236.99.252
connected: BLA_W1345096_0199 at IP 149.236.99.253
connected: DRU1 at IP 149.236.99.89
connected: ELCB_Z100818_2471 at IP 149.236.99.20
connected: SampleJet at IP 149.236.99.55
connected: IPSO at IP 149.236.99.90
read configuration data from BSMS/2
Connecting to ipsoserver at IP 149.236.99.90 ... done
IPSO: connected to spectrometer subnet.
configure AQS racks
read configuration from AQS
read BIS from AQS_RX1 ... done
read all RG values from AQS_RX1 ... done
read BIS from AQS_SGU1 ... done
read BIS from AQS_SGU2 ... done
read BIS from AQS_SGU3 ... done
read BIS from AQS_SGU4 ... done
edit parameters
send user input to server
wait for server to handle parameters
get permission to continue configuration
continue configuration
parse input from user
check for questions from server
edit parameters
send user input to server
wait for server to handle parameters
finish configuration on server
wait for server to finish hardware configuration
get permission to continue configuration
continue configuration
parse input from user
check for questions from server
get permission to finish configuration
finish configuration
configure remaining units
check HPPR preamplifier configuration
read HPPR/2 controller configuration
read preamplifier module configuration
read BIS from HPPR/2 module C1 ... done
read BIS from HPPR/2 module P1 ... done
read BIS from HPPR/2 module P2 ... done
read BIS from HPPR/2 module P3 ... done
read BIS from HPPR/2 module P4 ... done
read BIS from HPPR/2 module P5 ... done
connected to Bruker Automatic Changer
detect wiring and connections
detect RF wiring between SGUs and amplifiers:
wake up AQS rack 1 ... done
wait 1 second for boards in AQS rack to boot ... done
check if a transmitter detects a 390 MHz signal from SGU1-NORM ... done
check if a transmitter detects a 390 MHz signal from SGU1-AUX ... done
check if a transmitter detects a 390 MHz signal from SGU2-NORM ... done
check if a transmitter detects a 390 MHz signal from SGU2-AUX ... done
check if a transmitter detects a 186 MHz signal from SGU2-AUX ... done
check if a transmitter detects a 61 MHz signal from SGU2-AUX ... done
check if a transmitter detects a 92 MHz signal from SGU2-AUX ... done
check if a transmitter detects a 600 MHz signal from SGU2-AUX ... done
check if a transmitter detects a 390 MHz signal from SGU3-NORM ... done
check if a transmitter detects a 390 MHz signal from SGU3-AUX ... done
check if a transmitter detects a 390 MHz signal from SGU4-NORM ... done
check if a transmitter detects a 390 MHz signal from SGU4-AUX ... done
check if a transmitter detects a 186 MHz signal from SGU4-AUX ... done
check if a transmitter detects a 61 MHz signal from SGU4-AUX ... done
check if a transmitter detects a 92 MHz signal from SGU4-AUX ... done
check if a transmitter detects a 600 MHz signal from SGU4-AUX ... done
check if a transmitter detects blanking signal 1 of SGU1 ... done
check if a transmitter detects blanking signal 2 of SGU1 ... done
check if a transmitter detects blanking signal 3 of SGU1 ... done
put AQS rack 1 asleep ... done
detection of RF wiring between SGUs and amplifiers finished.
write configuration parameter file
write configuration information file
copy configuration parameters into diskless distribution
clear cache of AqRackServer
show configuration
hardware detection finished
allow other clients to start a configuration
execute edsp cf
additional setups